E N G L I S H
35
Fig. 8-58 Pop open clamps (Step 3)
Fig. 8-59 Manifold, showing position of the 3 solenoids (Step 8)
8.19 Replacing Inlet/Outlet Manifold
Tools and supplies required:
• 9/16” nut driver
• Phillips head screwdriver
• flat blade screwdriver
1. Remove bolts as in Step 8.18.2.
2. Remove the clamps as in Step 8.18.3.
3. Using Phillips head screwdriver, disconnect pressure transducer
from the Manifold.
4. Disconnect the entire Manifold Harness.
5. Remove the solenoids and valve stems using flat blade
screwdriver.
6. Remove the thermistor.
7. When reinstalling, connect the valve stems first, then the
solenoids, then the pressure transducer, then the thermistor.
8. When reinstalling the Manifold Harness, note that there are
labels on the harness identifying the solenoids (FV, BV, VV).
If the solenoids are not in the proper position as shown, the
device will not work properly (Fig. 8-59).
Fig. 8-60 Manifold Assembly
8.20 Replacing Level Sensor
Tools and supplies required:
• wire cutters
1. Disconnect gray cable from the I/O board (See Fig. 8-15,
I/O Circuit Card connections).
2. Using wire cutters, remove the cable tie
3. Remove the bracket holding the level sensor into the tank.
Fig. 8-61 Level Sensor
Fig. 8-62 Position of Level Sensor is indicated
FV
BV
VV