Patient Temperature 2 Probe Out
of Range
Patient Temperature 1 probe is not detected, or the temperature
reading is above the upper limit of the display range (44°C/111.2°F)
while in a patient control mode (e.g. Control Patient, Cool Patient or
Rewarm Patient).
29
Patient Temperature 2 Calibration
Error
The system is unable to internally check the calibration of the Patient
Temperature 2 channel within ± 1.0°C when the system is in a patient
control mode (e.g. Control Patient, Cool Patient or Rewarm Patient).
Patient Temperature 2 channel is made inactive.
30
Patient Temperature 2 Calibration
Error
The system is unable to calibrate the Patient Temperature 2 channel
within ± 1.0°C when the system is in Manual Control mode.
Patient Temperature 2 channel is made inactive.
31
Patient Temperature 2 Calibration
Error
The system is unable to calibrate the Patient Temperature 2
channel within ± 1.0°C when the system is in a patient control mode
(e.g. Control Patient, Cool Patient or Rewarm Patient).
Patient Temperature 2 channel is made inactive.
32
Patient Temperature 2 Calibration
Error
The system is unable to calibrate the Patient Temperature 2 channel
within ± 1.0°C when in Manual Control mode.
Patient Temperature 2 channel is made inactive.
33
Water Temperature High
The primary outlet water temperature is above 44°C (111.2°F).
34
Water Temperature High
The primary outlet water temperature is above 42.5°C (108.5°F)
35
Water Temperature Low
The primary outlet water temperature is below 3.5°C (38.3°F).
36
Water Temperature High
The secondary outlet water temperature is above 44°C (111.2°F).
37
Water Temperature High
The secondary outlet water temperature is above 43°C (109.4°F).
38
Water Temperature Low
The secondary outlet water temperature is below 3.0°C (37.4°F)
40
Unable to Maintain Stable Water
Temperature
In Manual Control mode, the system is unable to control the water
temperature within 1.0°C/°F of the water target after 25 min. in the
current mode or since last change to water target temperature.
41
Low Internal Flow
Insufficient internal flow during system priming or pre-conditioning.
43
User Settings Not Saved
The user settings are invalid and are unable to be saved. The saved
system default settings are restored.
44
Invalid System Log Entry
One or more of the entries into the system event log is invalid.
The system event log is used by clinical engineering personnel for
product service.
This issue does not affect the performance of the system to deliver
patient therapy.
45
AC Power Lost
The AC power was lost while the power switch was in the On position.
46
Control Panel Communication
The control panel is not communicating with the system.
47
Control Panel Communication
The control panel is not communicating with the system.
48
Patient Temperature Out Invalid
The Patient Temperature Out calibration data in non-volatile memory
is invalid.
50
Patient Temperature 1 Erratic
Patient Temperature 1 discontinuity. There has been a significant
change in patient temperature over the past 8 minutes.
51
Patient Temperature 1 Below
Control Range
Patient Temperature 1 is less than 31°C (87.8°F) while in a patient
control mode (e.g. Control Patient, Cool Patient, or Rewarm Patient).