A S 5 0 0 0 S E R V I C E M A N U A L
46
Due to the size and weight of the
A
rctic
S
un
®
5000 Temperature
Management System, it should be shipped on a pallet using
Medivance-provided packaging materials. If the original packaging is
not available, a shipping kit may be ordered from Medivance.
1) Perform a total drain of the system.
a) After the device has been drained, power On, from the
Therapy Selection screen, press the Advanced Setup button
to display the Advanced Setup screen.
b) Press the Start button next to Total Drain and follow the
instructions
2) Place the
A
rctic
S
un
®
5000 Temperature Management System
onto the white foam attached to the pallet and center the unit so
that it straddles the foam.
3) Place the square cardboard piece with the foam down on top of
the unit and place any accessories on top of the foam piece.
4) Slide the cardboard tube over the unit ensuring it meets flush with
the top surface of the pallet.
5) Using the strap provided, tightly secure the unit to the pallet.
Please firmly tighten the strap so the unit and its contents are
secured to the pallet for shipment.
Appendix F - Shipping