E N G L I S H
25
8.7 Removing Internal Components from Chiller Frame
Tools and supplies required:
• 7/16” nut driver
• small-blade flat head screwdriver
1. Remove six bolts holding internal components onto frame.
2. From the right side of the Control Module, carefully disconnect
the Chiller power connection gray cable.
3. If device is equipped with an AC pump from the left side of the
Control Module, disconnect the black compressor-evaporator
tubing from the white plastic fitting it connects to. Use the small
flat blade screwdriver to pop the snap fitting open. (Use pliers to
reclose the snap fitting upon reassembly.) If device is equipped
with a DC pump, from the left side of the control module,
remove black molded tubing connecting chiller pump to chiller
evaporator. Use a small flat-head screwdriver to loosen and
disconnect two clamps. Discard them.
4. From the front of the Control Module, place one hand
underneath the internal components and tilt them forward,
and then lift out the internal components.
Fig. 8-19 Remove six bolts (Step 1)
Fig. 8-20 Disconnect Chiller power connection gray cable;
location indicated (Step 2)
Fig. 8-21 Open the snap fitting (Step 3 - AC pump)
Fig. 8-22 Disconnect two clamps (Step 3- DC pump)