Normative Reference (Appendix)
Guidance and manufacturer’s declaration – electromagnetic immunity
The model LT 470 (“Mini Daylight” ) is intended for use in the electromagnetic
environment specified below. The customer or the user of the model LT 470
(“Mini Daylight” ) should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environ-
ment – guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and
voltage varia-
tions
on power sup-
ply
input lines
IEC 61000-
4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial
or hospital environment. If
the user of the model LT 470
(
“Mini Daylight”
) requires
continued operation during
power mains interruptions, it is
recommended that the model
LT 470 (
“Mini Daylight”
) be
powered from an uninterrupt-
ible power supply or a battery.
Power fre-
quency
(50/60 Hz)
magnetic field
IEC 61000-
4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical loca-
tion in a typical commercial or
hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for LT 470
Page 104: ...100 LOT IP20 1 GR...
Page 105: ...101 1 GR MEDISANA LT 470 S A D seasonal affective disorder...
Page 106: ...102 1 GR Normative Reference 1...
Page 107: ...103 1 GR 15...
Page 108: ...104 1 GR 10...
Page 110: ...106 3 2 1 On Off 1 I 2 On Off 1 0 15 35 cm 35 cm 2 25 cm 70 15 cm 30 3 GR 3 1 5 4 15 35 cm 3...
Page 111: ...107 3 4 GR 2 2 4 15 7 4 1 15 4...
Page 112: ...108 4 GR 4 3 4 2 15 4 6 4 4 1 4 5 8000...
Page 118: ...45222 West Ver 1 7 08 2016 MEDISANA AG Jagenbergstra e 19 41468 NEUSS Deutschland...