ECG Analysis Performance
Rhythm class
AHA-DB MIT-DB CU-DB
VF-DB
VT-DB Total number of
sample size
VF-shockable : TP
988
22
282
908
-
2200
VF-shockable : FN
44
0
7
23
-
74
VF-shockable : sensitivity(%) 95.74
100.00
97.58
97.53
-
96.75
Adult mode
VT-shockable : TP
-
-
-
701
-
701
Adult mode
VT-shockable : FN
-
-
-
66
-
66
Adult mode
VT-shockable : sensitivity(%)
-
-
-
91.40
-
91.40
Pediatric mode
VT-shockable : TP
-
-
-
-
18
18
Pediatric mode
VT-shockable : FN
-
-
-
-
3
3
Pediatric mode
VT-shockable : sensitivity(%)
-
-
-
-
81.82
81.82
Non-shockable : TN
32291
17518
1267
7062
-
58138
Non-shockable : FP
0
2
8
33
-
43
Specificity(%)
100.00
99.99
99.37
99.53
-
99.93
Positive Predictive Value
100.00
91.67
97.24
97.99
100.00 98.55
Database for ECG Analysis
From AHA (American Heart Association) official database
From MIT (Massachusetts institute Technology) official database
(CU-DB: The Creighton University Sustained Ventricular Arrhythmia Database,
VF-DB, VT-DB: MIT-BIH Malignant Ventricular Arrhythmia Database)
ECG rhythm to determine if a shock is appropriate
Ventricular Fibrillation at a amplitude greater than or equal to 0.2mV.
Ventricular Tachycardia at a heart rate greater than or equal to 150 bpm (Adult) / 180
bpm (Pediatric).