MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
42
42
Replacement of the pads must be carried out if:
The use by date of the pads has been exceeded.
When the pads have been used. (It is a single use item and must be replaced with
new pads.)
The package of new pads has been previously damaged.
The pads have been vent.
“
Replace Pads.
”
voice Prompt is played.
Pads Status icon is displayed even though the pads are connected.
Replacing Pads
1. Take the replacement pads from its protective bag.
2. Disconnect the pad connect from the AED.
3. Push the pads firmly to ensure it is fully inserted.
4. Turn on the AED.
5. Check the status indicator. If the pads have been inserted correctly, the status Indicator
displays
“
O
”
and the voice prompt
“
Unit ok.
”
is emitted.
6. If the status indicator displays
“
X
”
or the voice prompt,
“
Unit fail.
”
is emitted, inform
relevant safety officer or person responsible for maintenance of the AED.
7. Update the relevant information to show the date that the replacement of pads and
battery was placed into service.
8. Dispose of the old pads.
CPR Feedback Module Maintenance
Replacement of the CPR Feedback Module must be carried out if:
CPR Feedback is not performed despite the CPR Feedback Module connected to
the AED.
CPR Feedback from the CPR Feedback Module is abnormal.
Replacing CPR Feedback Module
Please refer to the Figure 7.