AED A16
Automated External Defibrillator
26
26
Replacing the CPR Feedback Module
Figure 7. Replacing the CPR Feedback Module
1. Loosen the CPR Feedback Module connector and separate the pads and CPR
Feedback Module.
2. Push the new CPR Feedback Module into the CPR Feedback Module connection
socket of the pads fully inserted and tighten the connection.
3. Keep the adhesive pads with the AED as preparation for emergency circumstances.
Note: When the exchange of the pads, connect the CPR Feedback Module that
connected to the pads which is needed exchange to the CPR Feedback Module
connection socket of the new pads.
Note: The CPR Feedback Module is not disposable, so that if the pads is needed to
exchange, connect the new pads and reuse.