MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
20
20
1 Hook
2 Battery
3 Battery release button
Figure 4. Inserting the battery
1. Pull down the battery while pushing the battery release button, then disconnect
the battery as shown in the Figure 3.
2. Prepare the new battery, then connect the battery and AED by using the hook as
shown in the second figure of the Figure 4.
3. When the connection part is fastened properly, slide the side of the battery with
the release button into the AED until the click sound is emitted as shown in the
third of the Figure 4.
The AED uses the non-rechargeable battery. The AED cannot operate with the battery that
has been completely discharged. Before turning on the AED with a battery that has been
completely discharged, first replace the battery. When the new battery is installed, the AED
is automatically turned on and then starts the battery insertion self test. After the battery
insertion self test is completed, the AED may then be powered off.
Figure 5. Removing the battery protection sticker
Note: For new batteries, remove the battery protection sticker before inserting the battery
as shown in the Figure 5.
1 2
3