30
30
material touching the patient during ECG analysis or defibrillation.
Note: Only pads supplied by Mediana should be used. Using other types of pads may
result in damage to the patient and the AED.
Note: When the voice prompt,
“
Replace pads.
”
, is played, ignoring the guidance and
using the same pads can cause the fail of the deliver the energy that AED assured.
Note: If the AED is turned on in out-of-range of environmental operation condition
described in this manual, temperature status will be blinked. In this case, place the
AED in appropriate temperature before use it.
1.
Check the status indicator displays “O”.
2. Remove the cover of the AED.
3. Turn on the AED by pushing the power button.
4. The AED automatically starts the Power-On-Self Test.
5. The test result is displayed on the status indicator and the voice prompt sounds.
Self test is passed : Voice prompt
“Unit ok”, Status indicator “O”
Self test is failed : Voice prompt
“Unit failed”, Status indicator ”X”
6. If the pads are inserted already, Adult mode is default. Whenever pushing the patient
mode switch button, user will hear the following voice prompt.
Adult icon
’
s LED lights up:
“
Adult mode
”
Pediatric icon
’
s LED lights up:
“
Pediatric mode
”
Figure 8. Patient mode switch Button
Note: The patient mode can switch even if any step except for CPR is going on
progress. If the patient mode is changed, the AED will emit the voice prompt
“Adult
mode
” or “Pediatric
mode
”. When the patient mode is changed du
ring CPR,
the AED will not emit the voice prompt.
Note: If the patient mode changes during Step 2 (ECG analysis) or Step 3 (delivering
electric shock), notify the user that the electric shock has been canceled, and
then repeat Step 2 (ECG analysis) again. However, if the electric shock is
canceled more than 3 times before the completion of Step 3 (delivering electric
shock), perform Step 4 (CPR).