31
2.
lb-oz mode
When the display locks on a weight.
Item
Header
Output data
(ASCII code)
Details
Control data
{0
Fixed at 16
Control data
~0
Fixed at 1
Model No.
MO
“xxxxxx”
6 byte fixed length
(“BWII”)
Body weight
Wo
xx.xx.x
5~7 bytes variable
length (unit: lb.oz)
Ex.) {0,16,~0,1,MO,”BWII”,Wo,11.14.2
When the Before Feeding key
1
is pressed to store weight.
Item
Header
Output data
(ASCII code)
Details
Before
Feeding
Ro
xx.xx.x
5~7 bytes variable
length (unit: lb.oz)
Ex.) Ro,11.14.2
When the After Feeding key
2
is pressed to store weight.
Item
Header
Output data
(ASCII code)
Details
Before
Feeding
Ro
xx.xx.x
5~7 bytes variable
length (unit: lb.oz)
After
Feeding
Po
xx.xx.x
5~7 bytes variable
length (unit: lb.oz)
Amount Fed
Go
xxx.xx.x
5~8 bytes variable
length (unit: lb.oz)
Ex.) Ro, 11.14.2, Po, 12.2.1, Go, 0.3.9
NOTE 1
Commas (,) are used to separate data.
NOTE 2
Terminator (end of data) is CR (ASCII code: 0DH), LF (ASCII code: 0AH).