WARNING
: Hospital waste. Treat the
following objects as hospital waste:
•
Inserts, when worn or broken;
• Inserts tightening wrench, when worn or
broken.
Disposable materials and materials that carry
biological risk must be disposed of according
to local regulations in force regarding
hospital waste.
WARNING:
When handling the inserts,
pay particular attention to the sharp, pointed,
irregular parts to avoid any wounds or
injuries.
multipiezo must be disposed of and treated as
waste subject to separate collection.
Failure to comply with the previous points can
result in a penalty pursuant to the directive
on waste electrical and electronic equipment
(WEEE).
The purchaser is entitled to deliver the device
to be disposed of to the dealer who supplies
them with new equipment; at Mectron,
instructions for proper disposal are available.
11
TECHNICAL DATA
Device compliant with Regulation (EU)
2017/745
Class lla
Classification under the IEC/EN 60601-1
I
Parts applied: type B (insert)
IP 20 (device)
IP 22 (pedal model FS-05)
Essential performance
According to the standard IEC 80601-2-60
the device has no essential performance
Device for intermittent operation
55sec. ON - 30sec. OFF with irrigation
30sec. ON - 120sec. OFF without irrigation
Power Supply
100-240 v~ 50/60 Hz
Max. Power Consumption
90 VA
Fuses
Type 5 x 20 mm, T 2AL, 250v
Working Frequency
Automatic Scan
From 24 KHz to 36 KHz
Powers
”endo”
“perio/scaler”
“restorative”
“soft mode”