multipiezo
FILLING
AND
FLUSHING
THE
IRRIGATION CIRCUIT
(Ref. 2 inside cover)
The device is equipped with the “pump/
flush” key, which depending on the
operating mode allows running either
the PuMP or the FLuSH function.
The PuMP function can be used at the
beginning of the treatment, to prime the
entire irrigation line up to the insert, so
that the treatment can be started with
the necessary irrigation (See
CAUTION
: The FLuSH function
(see
must be used
after every patient treatment, before
starting the cleaning and sterilization
procedures (See
). Failure to carry out flushing of
the handpiece tubing will lead to salt
crystallisation that can seriously damage
the device
WARNING
: The “FLuSH” function
does not substitute for the cleaning and
sterilization procedures described in
carried out the “FLUSH” function, the
cleaning and sterilization procedures
must be followed meticulously in order
to prepare the device accessories ready
for the next patient and minimize any
risk of patient-to-patient contamination.
perio/scaler
perio/scaler
restorative
restorative