42
multipiezo
Rinse the insert(s) in demineralised water;
13
H
₂
O
Rinse the internal channel of the insert
with demineralised water at a pressure of
3.8 bar for at least 10 seconds so as to
eliminate any soil residue.
14
x10 sec.
H
₂
O
8.4.1.4 Inserts Torque Wrench
Place the torque wrench in a clean
container, in a horizontal position. Add
enough enzymatic detergent solution to
completely cover the torque wrench.
Leave the torque wrench to soak in
the enzymatic detergent solution for
10 minutes at ambient temperature
(
22 °C ±2 °C). This procedure reduces the
amount of blood, protein, and mucus on
the instrument(s);
15