26
multipiezo
6
FLUSH FUNCTION
The FLuSH function allows to run a flushing cycle on the irrigation circuit of the handpiece used
during the treatment, by following the steps described in this chapter.
CAUTION
:
FLUSH function.
The FLuSH
function must be used after every treatment,
before starting the cleaning and sterilization
procedures.
WARNING
: The “FLuSH” function
does not substitute for the cleaning and
sterilization procedure described in this
manual. After having carried out the “FLUSH”
function, the cleaning and sterilization
procedures must be followed meticulously in
order to prepare the device accessories ready
for the next patient and minimize any risk of
patient-to-patient contamination.
CAUTION
: Failure to carry out flushing
of the handpiece tubing will lead to salt
crystallisation that can seriously damage the
device.
Fill the tank with water, preferably
demineralised water. Connect the tank to
the device;
1
Position the handpiice, with or without
insert, above a container to contain
the liquid that will outflow during the
flushing cycle;
2