47
CLEANING AND STERILISATION
Dry the insert tips torque wrench using a
soft cloth with low fibre release;
3
Lubricate the insert tips torque wrench
with FDA approved
medical-grade
lubricants by spraying it directly onto
the peripheral contact surface inside the
torque wrench, as indicated in the figure.
CAUTION:
Do not use oil or silicone-
based lubricants.
After having applied the lubricant,
remove any excess oil using a clean lint-
free cloth.
4
8.7 Sterilisation
8.7.1 Preparation
Seal the scaler handpiece (without
inserts), scaler nose cone, and light pipe
individually and separately in disposable
sterilisation bags.
CAUTION:
Use FDA approved,
appropriately sized medical steam
sterilisation pouch compliant with
standard uNI EN ISO 11607-1.
1