![MBRAUN EasyLab Operating Manual Download Page 88](http://html1.mh-extra.com/html/mbraun/easylab/easylab_operating-manual_1752773088.webp)
Operating Manual
Operation instructions
Page13-10
Pin-No.
Contact
1
Supply Ground
2
Switching ON/OFF 24 V
3
Signal Ground
4
Not Connected
5
Live bit (H2O)
6
Signal 0 - +10 V
7
24 V
8
Supply Ground
13.4.5 Installation
The Moisture Sensor is mounted on an appropriate vacuum-tight NW40 flange by means of a
centring ring and a clamp. The plug connection to the control unit should not be made before
the whole box-system has been purged sufficiently with inert gas. The operation of the Sensor
as well as the display of the measured values is controlled from the operator panel.
13.4.6 Sensor Cleaning
CAUTION
Risk of damage to process materials!
The Moisture Analyser must be cleaned every 2000 hrs. A
reminder is given as a warning when this service work is due.
To achieve optimal moisture measurements the sensor is recommended to be closely
inspected within a period of three months.