Operating Manual
Operating Manual Page 3-3
If the
MBRAUN
system is only part of a larger complete unit, the requirements of the other
system components have to observed as well.
3.1.3 Installation and Commissioning
We recommend, that a
MBRAUN
technician installs the system and performs the first
commissioning. If the customer does the installation and first commissioning by himself, please
adhere to the following instructions.
3.1.4 Final Inspection
A final inspection and specification test is performed by a
MBRAUN
technician if the installation is
made by
MBRAUN
. If the customer does the installation and first commissioning by himself,
MBRAUN
recommends to perform a complete system and specification test prior to operating the
system.
3.2 Gases Used by the System
3.2.1 Overview
Pressure
Reducing Valve
to customers
from gas
Working Gas IN
Regeneration Gas
Vacuum-
-
exhaust system
source 200 bar
6 bar (0.6 MPa)
Exhaust
depressurized
permanent) (20 MPa)
EASY
LAB
from gas
Regeneration Gas IN
source 200 bar
0.3-0.4 bar (30 – 40 kPa)
permanent) (20 MPa)
KF40
40
KF40
pump
5-6.0 bar (0.5 – 0.6 MPa)
Purge Gas IN
to customers
exhaust system
Purge gas OUT
GLOVE BOX