![MBRAUN EasyLab Operating Manual Download Page 61](http://html1.mh-extra.com/html/mbraun/easylab/easylab_operating-manual_1752773061.webp)
Operating Manual
Operation instructions
Page10-3
Attention:
The cycle program must shut down
before regeneration and also purifier
inlet and outlet of the butterfly valve
need to manually shut down.
Please shut down circulation by
clicking circulation button when the
circulation is active, and then
regeneration button is effective,
meanwhile the VPG keeps running.
Manually close the two butterflies at
purifier in and out port (The valve
handle should be parallel to the
ground)
Start reg. program by clicking
regeneration button, then the indicator
light will flash and system will test.
Clicking the button again if the gas
vent is verified.
It shows that regeneration program is
running if the light of regeneration
button become red.
10.3.2 Perform of regeneration program
The paragraph shows the regeneration process .All of the steps are automated.