Operating Manual
Operation instructions
Page 3-5
3.3.3 Additional Purge Gas
Use
Purging and filling the system with the desired inert working gas
(when
commissioning for the first time and after servicing or repairs of
the system.)
Gas type*
Same as Working Gas
Purity
Medium purity (4.8 or better); from bottles or other gas supply
facilities.
Quantity
Approx. 10 - 12 m
3
/m
3
box volume for purging the system when
commissioning
the system for the first time or intermediately purging the system.
3.3.4 Regeneration Gas
3.3 Equipment for Connections
3.3.1 Preparation
Prior to delivery of the system, the user will receive an information sheet specifying the
necessary accessories required to make the connections. The following specifications are a
general overview.
3.3.2 Equipment for Working Gas Connections
Pressure Reducing Valve for Working Gas:
Use
Working gas pressure control system.
Pressure,
200 bar (20 MPa) primary, 6 bar (0.6 MPa) secondary,
Flow rate
with a flow rate of 250 l/min
Connectio
n
Ø 10 mm Swagelok fitting.
Type
Supply Piping for Working Gas: