background image

Entrée Bath Gen III

 

OPERATION MANUAL 

  
 
 
 
 

01-27-2022 
Page 8 of 18

 

450 East 32

nd

 Street PO Box 1022 (68026) 

Fremont NE 68025 
800-798-5867  Office: 402-564-5867   FAX: 402-563-9102 
www.mastercarebath.com

 

4.

 

Transfer Individual to the Transfer Chair 

a.

 

Push down on the Open Door Button until the 
door is fully opened and stops on its own.  

b.

 

Position  the  comfort  pad  group  in  the 
Transfer  Chair  as  follows;  two  rectangular 
pads parallel about three to four inches apart 
centered  in  the  individual  seating  area,  one 
rectangular pad horizontally on the seat back 
at  approximately  the  lumbar  area  of  the 
individual 

 

c.

 

Push  the  Transfer  Chair  to  the  resident  and 
position for a normal wheel chair transfer. 

d.

 

LOCK  the  brakes  by  stepping  down  on  the 
LOCK-ARM  TAB  located  on  the  back  of  the 
rear casters. 

Failure  to  lock  the  caster  brakes  before 
the  resident  is  transferred  into  the 
Transporter  could  result  in  injury  to  the 
operator or resident.  

e.

 

Transfer  the  patient  on  to  the  Transporter 
using proper nursing transfer techniques. 

f.

 

Position  the  security  belt  attached  to  the 
connecting knobs on the lift chair having one 
belt  crossed  diagonal  from  one  shoulder  to 
the  opposite  side  hip  and  another  belt 
connecting the knobs on each side of the lap. 

(See the Belting Diagram)

 

 

 

Comfort Pads Position 

Open Door 
Button 

Bar 

Security Knob 

Security Knob 

Security Knob 

Comfort Pads 

CASTER BREAK 

LOCK-ARM TAB 

CASTERS 

Summary of Contents for MB-68 AV-XL

Page 1: ...llow All Instructions Thoroughly Important safety instructions Read and follow all instructions thoroughly CAUTION The caution heading identifies important maintenance and operation information Failure to obey a caution warning may result in damage to the Entree Bath and may void the warranty ...

Page 2: ...s such as a hair dryer lamp telephone radio or television within 5 ft 1 5m of the hydromassage bathtub Proper electrical grounding must be performed by a licensed electrician following local state and federal codes Test the ground fault circuit interrupter protecting this appliance periodically in accordance with the manufacturer s instructions For permanently connected units a green colored termi...

Page 3: ...n the door It will override the drain water sensors Door Open Close Button To open door press and hold down the Door Open Button until the door comes to a complete stop To close the door press and hold down the transmitter close button until the door comes to a complete stop CAUTION Check for obstructions in the sensors like wet clothes foot etc CAUTION Door will not open if drain sensors detect w...

Page 4: ...on at 60 of full volume To increase or decrease the air volume press the air heat button and hold down to increase level or the air heat button to decrease level until the desired level is achieved The heat level is tied to the air volume and adjusted the same way BathAire SM Wave Mode To create a wave effect Press the wave button once for the long cycle a second time for the short cycle and a thi...

Page 5: ...EM Liquid Dispensing Controls These controls are used to add Bath Oil and Shampoo directly into the bath water Disinfectant Valve Dispenses disinfectant through the jets Turn valve to the right for disinfecting To the left for flushing DISINFECTANT VALVE DISINFECTANT BOTTLE LIQUID DISPENSIG CONTROLS Bath Oil Button Shampoo Button SHAMPOO BOTTLE BATH OIL BOTTLE Liquid Dispensing Controls Liquid Dis...

Page 6: ...exceed 110 If you are unable to adjust to the normal level of 100 to 105 remove the individual and report the problem to maintenance Do not use bath until the mixing valve is serviced to correct the problem Failure to heed this warning could result in a scalding injury to the individual c The Reservoir Fill will shut off automatically when the reservoir is full CONSOLE CONTROLS Mixing Valve Reserv...

Page 7: ... is desired prior to turning on the BathAire SM remove the Aromatherapy Cap that is placed under the Dispensing Box Access Door place selected scent cartridge or sachet in the Aromatherapy Cartridge Holder and securely close the access cap prior to turning on BathAire SM See Aromatherapy Section in this manual CAUTION Do NOT remove beads from cartridge or sachet Never place liquids in the Aromathe...

Page 8: ...dual c Push the Transfer Chair to the resident and position for a normal wheel chair transfer d LOCK the brakes by stepping down on the LOCK ARM TAB located on the back of the rear casters Failure to lock the caster brakes before the resident is transferred into the Transporter could result in injury to the operator or resident e Transfer the patient on to the Transporter using proper nursing tran...

Page 9: ...ITY BELT d The resident s weight will appear in the display automatically Make sure no one is touching or leaning on the chair including yourself during the weighting procedure e Record the weight Note If you accomplish the above steps every time you weigh someone you will help assure getting an accurate weight SCALE CONTROL PAD OPERATIONS ON ZERO Button Press before seating person to be weighted ...

Page 10: ...table safe See Warnings Below After adjusting temperature turn shower control valve off Close drain with drain open close lever Turn fill control valve on Continue monitoring the thermometer to maintain safe temperatures until tub or Reservoir is filled The Reservoir once filled will shut off If you do not want to fill the Reservoir turn fill control valve off once bath is filled e If manual fill ...

Page 11: ...therwise ignored 6 A Red LED associated with the pump operation will turn on when the pump operation is activated 7 See the maintenance instructions for the purpose and function of the Green LED Replacing One Liter Jugs Priming Operation a Remove the empty bottle of solutions and replace with MasterCare s Bath Oil or Shampoo Remove the cap and seal under the cap on the bottle of liquid b Insert th...

Page 12: ...on open the drain by rotating the Drain Handle Overflow to open position Allow the tub to drain completely before opening the door Note Sensors in the tub will prevent the Open Door button from functioning while there is water in the tub Make sure tub is completely empty and nothing is touching the sensor i e foot or wash cloth etc CAUTION If an emergency arises press simultaneously the Emergency ...

Page 13: ...u can before removing them from the tub by leaning the individual forward and towel dry their back side and inside of the tub area behind the individual Lift each leg and towel dry the underside of the legs top of the seat under legs and feet foot well area under feet 11 Move the Individual out of the Tub 12 Finish drying the individual 13 You are now ready to transfer the individual back to their...

Page 14: ... 10 minutes or as recommended by the disinfectant s manufacturer 10 Turn on the flush water turn knob left run water through the disinfecting lines in the dispensing box Turn off turn knob upright 11 Thoroughly rinse all cleaned components of the tub and transfer chair CAUTION Some cleaners disinfectants and floor strippers contain ingredients that are corrosive or abrasive These solutions or comp...

Page 15: ...nfectant Cleaner Bath Oil and Shampoo solutions used during the bathing procedures must be stored in a secure area not accessible by the patients It is recommended that a two person bathing team be established for each shift with trained backups Using the team approach with one member transferring the individuals to and from the bathing area and the other operating the bathing system increases the...

Page 16: ...rCare bathing systems are equipped with onboard GFCI s making them easy to access for your convenience In the event a power surge trips the GFCI it can be accessed and reset using the following procedures GFCI Ground Fault Circuit Interrupter detects and interrupts electric faults to insure safety to all individuals Reset Button trips and shuts down all power should a power surge outage faulty out...

Page 17: ...mperature Fill Reservoir Console And Or Tub Use Liquids As Desired Use Shower Wand With Shower Wand Control Valve Open Drain With Open Close Lever After Water Is Drained Open Tub Door With Push Buttons Transfer Individual Out Of The Entree Bath Return Individual To Room Or Desired Location Remove Comfort Pads And Place In Foot Well Use Disinfectant To Clean Tub Transfer Chair And Pads Rinse Disinf...

Page 18: ... Street PO Box 1022 68026 Fremont NE 68025 800 798 5867 Office 402 564 5867 FAX 402 563 9102 www mastercarebath com It is recommended that the operations wall chart be posted on the wall next to the control end of the tub for quick reference by the operator of the GEN III Tub ...

Reviews: