Rad-5/5v
Chapter 4: Operation
www.masimo.com
36
Masimo
Menu Level 4 - LED Brightness and Factory Defaults
Push the Mode/Enter button again to enter menu level 4.
Button
Setting
LED Display Brightness (4 levels)
Note:
All LED indicators are
illuminated while adjusting this
setting.
Use Up or Down Arrow
Keys to adjust parameter
to desired setting.
Restore Factory Defaults
Yes/No
Save User Identified Default
Settings (Password: next button,
up arrow, down arrow, next button)
Pressing the Mode/Enter button a fifth time, returns the Rad-5 to patient monitoring in the
Saturation/Pulse Rate Mode. Additionally, the Rad-5 will automatically return to patient
monitoring display from any menu level/setting after 10 seconds with no key presses.
Special Menu
This section gives an overview of the Rad-5 special menu selections available. To navigate
through the menus, use the Mode/Enter, Next, Up and Down keys located on the front panel
of the oximeter. The following sub-sections describe each menu item in more detail. The
oximeter has options that allow user configuration to suit specific needs. These selections
apply only to the Rad-5.
Sleep Mode Operation
The Rad-5 Sleep Mode allows the device to capture normal and abnormal data without
triggering alarms. This mode blanks out the device display and disables the alarms with the
exception of the Battery Level Indicator and Alarm Silenced Indicator, even after a power
cycle. The keypad is locked except for the power button. However, any single key press brings
the display back for 10-seconds. Upon power up, SLP mode is displayed along with a 10-
second display of parameters. Pressing and holding both the Mode Enter and Next keys for 3
seconds (select next (STD), Mode Enter) will put Rad-5 back into the special menu to exit.
CAUTION:
Alarms are disabled in this mode.
Standard and Sleep Mode
Turn the device on, then push and hold the Mode/Enter and Next buttons simultaneously for
3 seconds to enter the special menu level.