Rad-5/5v
Chapter 4: Operation
www.masimo.com
34
Masimo
Menu Level 2 - Averaging, Sensitivity, FastSat, and SmartTone
Push the Mode/Enter button again to enter menu level 2.
Button
Setting
Averaging
1
The signal averaging
time of this device can
be set to:
2, 4, 8, 10, 12, 14 or 16
seconds
Use Up or Down Arrow Keys to
adjust parameter to desired setting.
Note:
The parameter is set/selected
when the Mode/Enter and Next
buttons are pressed.
Use the Up Arrow key to turn on
SmartTone.
Sensitivity
2
Hi = Maximum
Nor = Normal
APO = APOD
FastSat
1
On, Off
SmartTone
3
On, Off
1
FastSat is automatically enabled in 2 and 4 second averaging.
2
Defaults to APOD and Normal only. High sensitivity will default to normal.
3
The SmartTone feature uses a proprietary algorithm that will provide pulse tones during
excessive motion and low perfusion conditions. The pulse tone is based off an averaged
pulse rate measurement from the proprietary algorithm and may not identify irregular heart
beat patterns when there is excessive artifact present.