Troubleshooting Chart
The following chart describes what to do if the Rad-5/5v system does not operate properly or
fails.
Problem
Possible Cause
Recommendation
Device Does not
Power On
Low battery
Check / replace battery.
Verify that the trending feature is off, as it
may deplete battery life at a faster rate than
normal.
Continuous
Speaker Tone
Internal Failure
Device requires service.
Press the Alarm Silence button.
If alarm continues to sound, power down the
device and remove batteries.
No Speaker Tone
Pulse Tone
set to "mute"
Alarm
Suspend
enabled
Press Up Arrow (Rad-5) or Alarm Volume
Adjust (Rad-5v).
Press Alarm Suspend button until Alarm
Suspend Indicator is no longer illuminated
or flashing. See
Alarm Silence
on page 40.
Buttons Do Not
Work When
Pressed
Internal Failure
Return for service. See
Return Procedure
on
page 60.