4-7
User Preferences
Setting the Automatic Shut Off Delay
Time
To prevent the battery pack from discharging while the
calibrator is not in use, you can program the BetaGauge
II to automatically shut off after a pre-defined period of
idle time. Idle time is defined as a period with no key
presses. You can also disable this feature if necessary.
Procedure
To set the automatic shut off delay time, do the
following:
1.
Use the
Menu
key to locate the Setup menu item
Off,
then press the corresponding function key. The
lower portion of the LCD will appear similar to the
example shown below:
2.
To disable the automatic shut off, press the
function key corresponding to the menu item
Dis
.
3.
To enable the automatic shut off, press the function
key corresponding to the menu item
Enab
.
4.
To set the delay time for the automatic shut off
feature, press the function key corresponding to the
menu item
Set
. The lower portion of the LCD will
appear similar to the example shown below:
Auto off Enabled
Dis.
Enab Set
Set Auto Off time
200
seconds
Summary of Contents for BetaGauge II
Page 1: ...BetaGauge II Pressure Calibrator User Guide ...
Page 8: ...viii Specifying Post Test Status Information 7 23 ...
Page 11: ...xi ...
Page 15: ...xiv ...
Page 27: ...General Description 1 12 ...
Page 35: ...Input Modules 2 8 ...
Page 141: ...Application Notes 6 56 ...
Page 165: ...Doc Cal Mode 7 24 ...
Page 183: ...Specifications B 8 ...
Page 189: ...Service Information D 4 ...
Page 197: ...Index I 8 ...