Basic Operations
3-12
Setting the Damping Factor for
Ch1/Ch2
The BetaGauge II allows you to enable a damping
feature to reduce the display fluctuations caused by a
noisy input signal (pressure or electrical).
NOTE:
The BetaGauge II automatically measures
(or reads) each module three times per second.
The following settings are available:
Damping
Setting
The BetaGauge II …
0.0 (Disabled)
Updates the displayed value with
every measurement (three times per
second).
2.0
Averages 7 measurements and updates
the displayed value approximately
once every 2 seconds.
5.0
Averages 16 measurements and
updates the displayed value
approximately once every 5 seconds.
10.0
Averages 31 measurements and
updates the displayed value
approximately once every 10 seconds.