3-11
Basic Operations
Zeroing the Milliamp Input
Use the following procedure to zero the milliamp input
on the DC module:
1.
Install the module in either module bay (channel)
of the BetaGauge II.
2.
Ensure that the BetaGauge II is on.
3.
Ensure that the module’s milliamp jacks are open
circuit. This is done by ensuring that no test leads
are attached to either the positive milliamp jack or
the common jack.
4.
Ensure that the BetaGauge II is displaying the first
four Setup menu items (
Trip
,
Unit
,
Damp
, and
V/mA
).
NOTE:
Only this menu display allows for both
channel selection and
ZERO
key operation.
5.
Press the channel select key corresponding to the
module bay where the DC module is installed
(
CH1
for the left side, or
CH2
for the right side).
6.
Press the
ZERO
key.
7.
The milliamps display should now be zeroed.
NOTE:
The milliamps display may drift slightly
from zero in the first few minutes after initial
power up. If so, press the
ZERO
key again while
the appropriate channel is still selected.
Summary of Contents for BetaGauge II
Page 1: ...BetaGauge II Pressure Calibrator User Guide ...
Page 8: ...viii Specifying Post Test Status Information 7 23 ...
Page 11: ...xi ...
Page 15: ...xiv ...
Page 27: ...General Description 1 12 ...
Page 35: ...Input Modules 2 8 ...
Page 141: ...Application Notes 6 56 ...
Page 165: ...Doc Cal Mode 7 24 ...
Page 183: ...Specifications B 8 ...
Page 189: ...Service Information D 4 ...
Page 197: ...Index I 8 ...