![Martel BetaGauge II User Manual Download Page 169](http://html1.mh-extra.com/html/martel/betagauge-ii/betagauge-ii_user-manual_1732374169.webp)
Battery Care/Maintenance
A-4
Battery Packs
The BetaGauge II uses rechargeable nickel–cadmium
(NiCd) batteries that have high power-to-size-and-
weight ratios. If properly maintained and cared for, they
will provide years of excellent service. However, NiCd
batteries are susceptible to damage under the following
conditions:
Heat.
Protect the batteries from prolonged
exposure to extremely hot temperatures.
Cold.
Protect the batteries from prolonged
exposure to extremely cold temperatures. If
possible, when batteries are cold, allow them to
warm to room temperature before using them.
Charging.
Charge batteries at an ambient
temperature of 0 to 45ºC.
Overcharging.
Overcharging NiCd batteries is the
most common cause of early failure. Do not charge
the battery packs continuously. For a full charge,
the trickle charger requires about 12 hours . (Never
charge the batteries for more than 16 hours.)
Approximately once a month, allow the batteries to
fully discharge before recharging.
Storage.
Store battery packs at an ambient
temperature of 0 to 35ºC. If you must store a
battery packs for a prolonged period (more than
three months), you should discharge the battery
pack prior to storage. The recommended method
for discharging a battery pack is through use in a
BetaGauge II calibrator.
Summary of Contents for BetaGauge II
Page 1: ...BetaGauge II Pressure Calibrator User Guide ...
Page 8: ...viii Specifying Post Test Status Information 7 23 ...
Page 11: ...xi ...
Page 15: ...xiv ...
Page 27: ...General Description 1 12 ...
Page 35: ...Input Modules 2 8 ...
Page 141: ...Application Notes 6 56 ...
Page 165: ...Doc Cal Mode 7 24 ...
Page 183: ...Specifications B 8 ...
Page 189: ...Service Information D 4 ...
Page 197: ...Index I 8 ...