3-13
Basic Operations
Reset Damping Guidelines
The following reset damping guidelines apply:
The BetaGauge II automatically disables (resets to
0.0 seconds) the damping feature for both channels
each time you turn the calibrator off and on.
The BetaGauge II automatically disables (resets to
0.0 seconds) the damping feature for a particular
channel each time you install, or re-install, a
module into the module bay corresponding to that
channel.
Procedure
To enable and set the BetaGauge II damping feature for
a specific module, do the following:
1.
Press the channel select key corresponding to the
module bay where the appropriate module is
installed (
CH1
for the left side, or
CH2
for the
right side).
2.
Use the
Menu
key to locate the Setup menu item
Damp
, then press the corresponding function key.
The lower portion of the LCD will appear similar
to the example shown below:
In the above example:
x = the channel number of the module.
CHx Damp Factor 0.0
0.0
2.0
5.0
10.0