Application Notes
6-50
SYS Mode
There are two methods for configuring a pressure
indicator or loop calibration using SYS mode. The
following table indicates which method to use.
Use this method if…
Method 1
Your field instrument has a pressure
input range within the ability of the
pressure module on the BetaGauge
II,
and
You are using either a local
or
remote pressure indicator to indicate
the pressure input signal to the field
instrument.
Method 2
Your field instrument has a pressure
input range that exceeds the ability
of the pressure module on the
BetaGauge II,
and
You are using either a local
or
remote pressure indicator to measure
the pressure output signal from the
field instrument.
Method 1
Software Instrument Setup
Pressure input
Pressure indicator output (requires manual
entry on the BetaGauge II)
Other Necessary Equipment
Pressure source