4-9-2 Hi, OK, Lo Comparison
1. Normal HOLD ( GEF-11 = 0 )
Entering the Hold mode, Hi, OK, Lo comparison output. Escape Hold mode will switch off the
outputs.
2. Peak HOLD ( GEF-11 = 1, 2 )
If Peak Ready is ON, Hi, OK, Lo comparison output. Escape Hold mode will switch off the
outputs.
3. The absolute value of peak HOLD ( GEF-11 = 3 )
Entering the Hold mode, Hi, OK, Lo will refer to Peak value to do the comparison.
4. GEF-11 = 4
When the external input single Judgement is ON, Hi, OK, Lo will refer to Peak value to do
the comparison.
4-10 Totalizing
(ACCU.) Auto / Transmit
With automatic totalising active (SQ-17) or RS232 / RS485 or BCD output set
to auto transmit.
1. SQ-01 = 1, 2, 4 or 5 batch / loss-in weight
a) When the weight reaches the Final weight and the batch finish signal is ON the net
weight will be added to the totaliser and number of additions is incremented. The
RS232 / RS485 and BCD outputs transmit data.
b) When the net weight returns to the zero range (GEF10), then the sequence in a) above
can be repeated.
2. SQ-01 = 3 Comparison mode
a) When the net weight exceeds the zero range and the weight has stabilized it will be
added to the totaliser and number of additions is incremented. The RS232 / RS485 and
BCD outputs transmit data.