3-6 Linearity Calibration
** Before the Linearity calibration, the General calibration should be completed.
Set CAL switch to ON
Select linearity calibration
Press the Key
Select one of five calibration, and
press the Key
Press the Key
When stable, the display area shows
the modified weight value
Please refer to the error message list if the display shows
.
X
Increment flashing
digit
Decrement flashing
digit
Move flashing point
left.
Move flashing point
right
Store data in
memory
Exit
/
Escape
.
. . . . .
Use to select one
of calibration points (1P~5P)
: no setting value
: with setting value
Press key
Press key
Key in the correct weight value
Press key
Press key
Finish the 1
st
calibration point
setting. Either continue the
second point calibration or
press key to exit the
linearity calibration process.
The current shows on the
screen and the indications
keep flashing