19
Item
Function
Setting value
Default
Parameter
Description
CSP-01
Unit
0
None
2
1
g
2
Kg
3
t
4
lb
CSP-02
Decimal
Point
0
None
0
1
1 Decimal Point
2
2 Decimal Point
3
3 Decimal Point
CSP-03
Division
1
Division size
1
2
5
10
20
50
CSP-04
Max.
Capacity
999999
000000
Max. capacity
999999
CSP-05 Zero range
0 =full range
(
1%~30%)
Zero range = calibration zero point
(Max.
capacity
setting value %)
0
CSP-06
Time of
zero
tracking
0.0 ~ 5.0
(sec)
Time and range of zero tracking should be use
at the same time. If the time is set to 0.0, the
zero tracking function is disabled.
1.0
CSP-07
Range of
zero
tracking
0 ~ 9
Range of zero tracking = (setting value
½
)D ,
D=min. division
Range and time of zero tracking should be use
at the same time. If the range is set to 0, the
zero tracking function is disabled.
2
CSP-08
Investigate
time in stable
0.0 ~ 5.0
(sec)
Investigate time and range should be use at
the same time. If the time is set to 0.0, the
investigate time is disabled.
1.0
CSP-09
Investigate
range in
stable
0 ~ 9
Investigate time and range should be use at
the same time. If the range is set to 0, the
investigate range is disabled.
2
CSP-10
Weight
unstable,
function
ZERO and
TARE
0
Action
0
1
None
CSP-11
Gross Weight
is negative,
function
TARE
0
Action
0
1
None