Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
44
General
Cleaning and disinfection
7
Cleaning and disinfection
7.1
General
All the components in the septic fluid suction
pump which come into contact with secretions
will have to be cleaned and disinfected after
each use.
7.2
Removing the septic fluid jar
⌧
Switch off the septic fluid suction pump.
⌧
Detach the extraction catheter and fingertip
from the suction tube(
Fig. 33) and dispose
of properly.
⌧
Detach the connection tube from the bacteria
filter.
⌧
Detach the suction tube from the patient-side
connector nipple.
⌧
Remove the septic fluid jar.
⌧
Remove the septic fluid jar cap.
⌧
Empty the septic fluid jar.
W
A R N I N G
!
Infection hazard!
Any and all the parts of the product could be contaminated.
Wear gloves and be absolutely sure to follow the hygiene rules during all cleaning
and reconditioning work.
C
A U T I O N
!
Property damage due to changes in materials!
Almost all the components in the product are made of plastic. Solvents, some dis-
infectants and some cleaning agents can soften plastic or cause tension fissures.
Do not use alcohol-containing agents to clean the surfaces. Follow the instruc-
tions for using disinfectants.
N
O T E
If the product is used in the home setting then the hoses and the hydrophobic bac-
teria filter will have to be replaced and the septic fluid jar and its cap will have to
be sterilised or replaced with each change of patients.