1
Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
16
VENTA MC 16 septic fluid suction pump
Introduction
1.4.10.4
Suction tube
The suction tube makes the connection be-
tween the connector nipple on the patient side
and the fingertip or the utensil.
Technical specifications
• Shore hardness 60
• Inside diameter 6 - 8 mm
• Length 1.3 m - 3.0 m
• Vacuum resistant down to -95 kPa
Requirements
• The suction tube shall comply with the hospi-
tal's standards for hygiene.
• The vacuum hose may not collapse.
• The outside diameter of the connector nipple
on the patient side at the septic fluid jar cap
shall match the inside diameter of the suction
tube.
1.4.10.5
Fingertip
The fingertip serves to vent the suction tube in
order to quickly be able to interrupt the extrac-
tion procedure.
Requirements
• It must be possible to sterilise the fingertip or
it must be a disposable item.
• The outside diameter of the connector nipple
on the patient side shall match the inside di-
ameter of the suction tube.
1.4.10.6
Utensil
The extraction catheter, lance etc. is designat-
ed as the utensil. It is used to extract the secre-
tion.
Requirements
• The inside diameter of the utensil's connec-
tor must match the outside diameter of the
fingertip.
• It must be possible to sterilise the utensil or it
must be a disposable item.
• Biocompatibility
1.4.10.7
Equipment carrier interface
The equipment carrier interfaces are used to
accept a rinsing jars or septic fluid jar and a
holder for rinsing jars or septic fluid jars.
Requirements
• Maximum load on the equipment carrier in-
terface is 1 kg.
• The interface for the rinsing jar or septic fluid
jar or for the holder for the rinsing jar or septic
fluid jar shall correspond to the equipment
carrier mount as per DIN EN 12218.
1.4.10.8
Rinsing jar
Any container meeting the following require-
ments may be used as a rinsing jar.
Requirements
• The rinsing jar shall have a volume of at least
250 ml.
• The rinsing jar shall be easy to clean and dis-
infect.