Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
29
First-time use
4
Mounting accessories
4.8
Mounting accessories
4.8.1
Mount septic fluid suction pump on trolley
A trolley is available to facilitate moving the
septic fluid suction pump and installing addi-
tional accessories.
Fig. 20: Engaging castor brakes
Engaging castor brakes
⌧
Engage the brakes (2) at the castors on the
trolley (1).
Fig. 21: Attaching the septic fluid suction pump
Attach the septic fluid suction pump
⌧
Place the septic fluid suction pump (3) on the
trolley.
The septic fluid suction pump is positioned
on the locating pin (4).
⌧
Use the star knob to turn down the screw (5)
on the trolley and secure the septic fluid suc-
tion pump.
⌧
Check to ensure that the septic fluid suction
pump is firmly seated.
W
A R N I N G
!
Injury hazard!
The septic fluid suction pump and its accessories may be mounted only on the
trolley (REF 5752 2356) designed for this purpose.
2
1
3
4