Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
33
Operation
5
Functional check
5
Operation
5.1
Functional check
Prior to using the system, the operator shall
confirm that the product is fully functional and in
good condition.
Check before each use to ensure that:
• The mains cable is undamaged.
• Components made of plastic or rubber (e.g.
control panel film, tube, septic fluid jar cap,
septic fluid jar) are in good condition and
show no damage due to aging.
• The overflow protection device and hydro-
phobic bacterial filter are mounted and func-
tional.
• The overflow protection device and hydro-
phobic bacterial filter have been properly
cleaned and neither residues nor grime is
present.
• Tube connectors and septic fluid jar cap are
tightly seated and do not leak.
• That no mechanical forces are applied to the
tubes.
• Tubes may not be kinked.
• Maximum vacuum of approx. -80 kPa is
reached within about 20 seconds when the
connection tube is held shut.
• The vacuum can be regulated steplessly
throughout the entire range.
• The septic fluid jar is attached to the septic
fluid suction pump.
• Septic fluid suction pump has been properly
cleaned and neither residues nor grime is
present.
• Damaged parts have been replaced by new
parts.
5.2
Extraction
W
A R N I N G
!
Health hazard!
The septic fluid suction pump is used in patient treatment. Any restriction of per-
formance can result in complications during treatment.
Check the unit on proper functioning every time before using and preparing it.
N
O T E
Connecting several septic fluid jars in series can cause delayed suction effect and
reduced suction power.
W
A R N I N G
!
Infection hazard resulting from using the wrong hydrophobic bacteria filter!
The septic fluid suction pump may be used only with a hydrophobic bacteria filter
suitable for the specific application. This is to protect it against contaminants
which might be present in the form of particles or aerosols in the gas being drawn
in and to prevent contamination of the air nearby.
Always use a hydrophobic bacteria filter appropriate to the particular extraction
situation.