Malfunctions and troubleshooting
6
Replace mains fuses
6.2
Replace mains fuses
Fig. 37: Replacing fuses
⌧
Disconnect the equipment plug.
⌧
Use a small screwdriver to carefully release
the catch (1) at the centre of the equipment
socket (2) and remove the fuse carrier (3).
⌧
Remove the fuses (4) from the fuse carrier.
⌧
Insert new fuses.
⌧
Insert the fuse carrier.
The catch will engage with an audible
click.
⌧
Ensure that the fuse carrier is properly seat-
ed.
W
A R N I N G
!
Electrical shock!
Disconnect the electrical plug before changing the line fuses.
C
A U T I O N
!
Property damage!
Use only fuses of the following type:
2 x T 1.25 A/H; 250 V; 5 x 20 mm
1
2
3
4