7
Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
48
Disinfection
Cleaning and disinfection
7.4
Disinfection
W
A R N I N G
!
Injury hazard!
Disinfectants may contain substances which are detrimental to health and could
cause injuries if they came into contact with the skin or eyes.
Protect the skin and eyes and observe the hygiene rules when working with dis-
infectants.
W
A R N I N G
!
Functional failure!
Check the unit's components for proper functioning after each cleaning or disin-
fection procedure.
C
A U T I O N
!
Property damage resulting from improper handling!
Be absolutely sure to follow the instructions for use provided by the manufacturer
of the disinfectant.
C
A U T I O N
!
Property damage at temperatures above 50°C!
The septic fluid jar may be used again only after it has cooled down.
C
A U T I O N
!
Property damage due to changes in materials!
The natural aging of plastics will be accelerated by autoclaving using live steam.
The functioning of the unit's components can be adversely affected by changes
in the materials.
Check the components for proper functioning after autoclaving.
N
O T E
The septic fluid jar, all the components in its cap, and the tubes are consumable
items. Depending on the cleaning process used, they will be subjected to a great-
er or lesser amount of wear and tear, this being due to the materials employed.
Inspect all components for serviceability before use. Replace them if there are
any signs of damage.
N
O T E
Using drapes which are not colourfast can cause discolouration in plastic compo-
nents.