Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
39
Operation
5
Empty the septic fluid jar
Fig. 32: Detaching the hydrophobic bacteria filter
Disconnect the bacteria filter and tubes
⌧
Switch off the septic fluid suction pump.
⌧
Detach the connection tube(2) from the hy-
drophobic bacteria filter (3).
⌧
Detach the suction tube (4) from the patient-
side connector nipple (5).
⌧
Remove the hydrophobic bacteria filter (3)
from the septic fluid jar cap (6).
Fig. 33: Removing the septic fluid jar.
Remove the septic fluid jar.
⌧
Press against the septic fluid jar (7) from the
front.
⌧
Carefully slide the septic fluid jar (8) out to-
ward the rear.
Fig. 34: Removing the septic fluid jar cap
Remove the septic fluid jar cap
⌧
Carefully grip the septic fluid jar cap (9) near
the patient-side connector nipple (10) and lift
the cap off the septic fluid jar (11).
⌧
Dispose of the extracted fluid.
2
3
5
4
6
8
7
10
9
11