OPERATING INSTRUCTIONS

 

ME3000

 

 

  Switch the power ON to unit. In a few seconds, the display shows SET and RDY(ready), with a zero quantity for RDY,  all LED 

indicators and alarms are off. The unit is ready for batching or configuration. 

 

  BATCHING: When SET/RDY is displayed, entering a batch quantity via the numeric keypad buttons will cause the displayed SET  

value to change and flash. Then, pressing the SET button locks in the new value, or pressing CANCEL reverts the value to its 
original setting, and the display digits stop flashing. 

  Press the START BATCH button to start batch batching the set quantity. The RDY line title changes to REC (received). 

  The voltage contact drive activates, the RUN LED illuminates indicating pump or solenoid are energized, followed by FLOW LED 

illuminating, indicating pulsing and operation of flowmeter. The digits on the REC (received) display line begin counting upward 
towards the selected batch quantity. 

  Upon REC digits reaching the selected batch quantity the alarm sounds (short beep) indicating completion of batch; RUN, FLOW 

and OUTPUT LEDs turn off. The displayed SET and REC values should correspond. If REC digits overshoot target, then scale back 
the difference by changing the Preact value via the Configuration Mode (see below). 

  To interrupt batch, push STOP button; REC counting will stop, and drive contact goes off. Push START BATCH to resume batch. 

  On batch completion or termination, press RESET. Display REC line title changes to RDY (ready).  

  To dispense the same set quantity again, press START BATCH. Otherwise, use the keypad to enter a new batch quantity. 

Warning: if RUN or FLOW LED indicators are on, but controller is not counting, discontinue use and call for  service. 

 

  EVENT LOG: using a straight-through female- to-female DB9 cable, connect from the Batch Controller’s (optional) RS232 port to a 

Laptop/PC serial port. On the Laptop, start the HyperTerminal software (supplied as part of Windows), set up a connection at 9600 
baud, 8 Data Bits, no parity, 1 stop bit, and capture text to a file. On the Controller, press the CANCEL and RESET buttons (those 
marked with ‘#’) simultaneously for 5 secs, and the Event Log will download. To stop download, press CANCEL for 2 secs. 

 

  CONFIGURATION: The Controller is factory configured to your requirements, but can be re-programmed by the user if required. 

  Configuration Mode is entered from the SET/RDY display by pressing the CANCEL and STOP buttons (those marked with a ‘*’) 

simultaneously for 5 seconds. 

  Use the buttons with the up/down arrows to step through configuration settings.  

  When a configuration value is diplayed, entering a new value via the keypad causes the display value to change and flash. Then, 

pressing SET locks in the new value, or pressing CANCEL reverts value to its original setting, and the display digits stop flashing. 

  Use the up/down arrows buttons to step through more configuration settings, or press CANCEL for 2 seconds to exit to the 

SET/RDY display. 

CALIBRATION   

1) The Batch Controller is set up for the connected flowmeter using the Calibration Input pulse scaling item under Configuration Mode, 
to match the flowmeter’s output pulse value.  

On-site calibration adjustment and test:

 

2)  Must adjust what is shown as received (REC) on the Batch Controller LCD display to match a known amount dispensed. So, set 

batch quantity  to say 190L, and batch into a 200 litre (44 gallon) drum. 

3)  If the amount collected is 

more

 than the REC amount shown on the LCD display, then 

decrease

 

the

 

Calibration Input

 value 

by the same % difference  e.g. if collected 200L when 190L on LCD, this is 10L more or 5% over (10/190x100%).  
So, decrease the calibration value by 5% i.e. if Calibration Input value is 300, new value is 300-5% = 300-15 = 285. 

4)  If the amount collected is 

less

 than the REC amount shown

 

on the LCD display, then

 increase the Calibration Input value 

by the 

same % difference. e.g. if collected 180L when 190L on LCD, this is 10L less or 5% under (10/190x100%).  
So, increase the calibration value by 5% i.e.  if Calibration Input value is 300, new value is 300+5% = 300+15 = 315. 

5) 

PREACT: 

To

 

calibrate  inflight  overflow,  enter  Configuration  Mode  (as  described  above),  and  set  the  Preact  value  to  the  same 

overflow reading as indicated by the LCD display, where overflow = (the REC quantity) - (the SET quantity). 

 

Example: You set 190 Litres, batch the quantity, 200 Litres is shown on display, and 200 Litres is collected in drum.  A valve may take 
extra time to close, so the received quantity usually overshoots. So, set 10 Litres on Preact to deduct the 10 Litres overshoot.  Next 
batch, the set batch quantity, LCD REC reading and amount collected in drum are all 190 Litres.

 

 

SPECIFICATIONS 

 

 

shown with optional 

 -SC and -5P interfaces 

Power supply 

220-250 vac  (optional 110 vac or 12-24 VDC) 

Fuse 

1 Amp (5 x 20mm case) 

Frequency input 

5 KHz 

Event Log 

internally records up to 300 batches 

Output to flowmeter 

12 VDC upto 100mA 

Relay 

Max. 240 vac, 1 A. Open Contact on request. 

Display   

2 line x 16 character, for quantity set and dispensed. 

Connection 

10 pin Weidmuller mating plug and socket 

Batch entry 

quantity selection and commands via IP64 keypad 

Optional Outputs 

scaled pulse output; 4-20mA output 

Optional interfaces 

RS232; PLC/Computer stop/start/reset 

Instrument housing 

ABS hi-impact case mould 

Mounting 

Panel mount. Panel cutout :190 L, 122 H mm 

External dimensions 

206 L, 130 H, 90 D mm.  

Weight   

1 kg 

 

Due to continuous product improvement, specifications are subject to change without notice. 

Summary of Contents for ManuFlo ME3000-RPFS

Page 1: ...flowmeter as per the installation guide found on the flowmeter brochure Install the Rota Pulse paddlewheel sensor RPFS P in a straight pipe section with the same diameter pipe as the adaptor tee section with 10x diameter before and 5x diameter after the sensor with no elbows reducers valves or restrictions within this pipe run Where the sensor is housed the pipe must be full when measuring D Use s...

Page 2: ...nd continuously if an error occurs With the ME3000 Batch Controller using the same instrument housing and the same 10 pin Weidmuller receptacle plug as other ManuFlo Batch Controller models changeover or upgrade is instant with no rewiring necessary It can be easily interfaced with PLCs through the optional computer control interface thus incorporating the controller s safety features and providin...

Page 3: ...nfiguration value is diplayed entering a new value via the keypad causes the display value to change and flash Then pressing SET locks in the new value or pressing CANCEL reverts value to its original setting and the display digits stop flashing Use the up down arrows buttons to step through more configuration settings or press CANCEL for 2 seconds to exit to the SET RDY display CALIBRATION 1 The ...

Page 4: ...nel Eagle etc PLCs MC 4 pin PLC Computer Command Start Stop Reset interface plug MC2 2 pin plug for scaled open collector pulse output Includes 4 pin external command Start Stop Reset interface plug e g ME3000 is the standard Batch Controller 240vac powered without any of the options whereas ME3000 MC2 is an ME3000 Batch Controller with a scaled open collector pulse o p and external Start Stop Res...

Page 5: ...tion to larger size pipes is possible depending on pipe wall thickness alternatively the Long Stem LS versions with adaptors are then used With only one moving part and limited intrusion into the pipe and combined with its flow through design the RPFS allows accurate measurement of liquid flows with virtually no head losses Each of the 4 blades of the rotor paddlewheel extends approximately one ce...

Page 6: ...G 6 1 Unscrew the black PVC locking cap anti clockwise 2 Hold the neck of the Tee piece in your left hand grasp the RPFS body with your right hand and turn slowly anti clockwise until the sensor hydraulically raises out of slot then pull upwards out of the socket never by the cable When returning the sensor to nipple adaptor insert so the keyway and slots line up then then push down until they loc...

Page 7: ...e If mounted in horizontal or inclined pipe make sure insertion position of sensor is at top or 45 from top not on the underside Selection of pipe diameter For best results use the table below Pipe size Flowrange Litres min Pulses Litre mm Min Max approx 1 2 20 13 160 116 25 20 250 75 32 30 410 46 40 50 640 30 50 90 1000 20 63 132 1580 11 7 65 120 1690 12 75 180 2250 8 3 80 190 2560 7 3 90 244 324...

Page 8: ...25 LS 140 mm SC140 LS 150 mm PVC150SC LS SCP150 LS SC160 LS 160 mm SCPE160 LS SC160 LS 200 mm PVC200SC LS SCP200 LS SC200 LS 225 mm SCPE225 LS SC225 LS 250 mm SCP250 LS SC250 LS 280 mm SC280 LS 300 mm PVC300SC LS SCPE300 LS SC315 LS Galvanised iron threaded ends BSP female 2000 kPa Note 25mm can be supplied with straight pipe sections already fitted Part GAL25 T2 PVC T piece Class 18 Cat 19 Glue i...

Page 9: ...4 ME3000 ME995 wiring with RPFS or other powered flowmeters ...

Page 10: ...5 ME3000 ME995 wiring with RPFS or other powered flowmeters ...

Page 11: ...6 PLUMBING INSTALL options guide for RPFS or other ManuFlo flowmeters ...

Page 12: ...7 ...

Page 13: ...blem Calcium buildup on wheel soak in diluted acid Warn bushes Backflow alarm Non return valve faulty jammed open Clean service or replace after batch complete Solenoid valve not properly closed damaged seal faulty solenoid Batch target display counter Flowrate too fast excessive overflow Turn down gate valve to restrict flowrate counts past batch selection or set preact overflow deduct function t...