The measuring range in which the calibration is to be performed is shown in brackets next
to the current measuring range. In Fig. 75, the current measuring range is “4” and the mea-
suring range of the calibration is “1”.
The test gas concentration, here “0.00”, is shown above the green arrow on the right hand-
side at the bottom of the screen. The green arrow is pointing at a calibration range of “-5.0
to +5.0 vol%”. The value of the actual gas and the test gas concentration needs to be in this
predefined measuring range.
To adjust the calibration range or test gas concentration, please tap on the edit button.
The M4/S2 screen opens with the channel-specific parameter “3 = Calibration/Adjust-
ment”. Here you can enter the parameters of your test gas and change the calibration
range.
Fig. 71: Setting the channel-specific calibration parameters
Please set the values to meet your calibration requirements. After adjusting the parame-
ters, please tap on the Calibration button M5 to confirm your entries.
Example of a manual calibration procedure with span gas
Fig. 72: Manual calibration with span gas
Example: in “'Fig. 72 Manual calibration with span gas'” the test gas has a 20.93 vol% oxy-
gen concentration.
Summary of Contents for GenTwo PMA1000V2.4
Page 90: ...89 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 91: ...90 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 93: ...92 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 94: ...93 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 95: ...94 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 96: ...95 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 97: ...96 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 98: ...97 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 99: ...98 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com...
Page 109: ...108 GenTwo PMA1000V2 4 1 00 00 www mc techgroup com Room for your own notes...