Malfunction
Possible cause
Possible solution
Drives cannot be
operated using the
hand controller
Mains cable is not
plugged in
Plug in the
mains cable
No power to socket
Check socket
Cable plug con-
nection not firmly
connected
Check plug connec-
tions on the motor
and hand controller
Hand controller or
drive is faulty
Inform the operator,
specialised dealer or
our customer service
Functions locked on
the hand controller
Enable functions on
the hand controller
(
)
Motorised adjustment
system not working
properly
Obstruction in the
adjustment area
Check moving parts
and remove any
obstructions
The safe working load
has been exceeded
Reduce the load
Drives cut out after
a long period of oper-
ation
The adjustment time
or safe working load
has been exceeded
and the control
unit has reacted to
overheating
Allow the drive
system to cool down
sufficiently
Opposite function
activates when
operating the hand
controller button
Motor plugs mixed up
Check that the
cables are connected
correctly or inform
your operator, spe-
cialist retailer or our
customer service
Side rails can no
longer be properly
adjusted
Mechanism blocked
or bent
Check all moving
parts and remove
any obstructions or
contact our customer
service
Castors do not brake
or cannot be rolled
Foreign objects have
become trapped in
the castors
Remove foreign
objects
The castor system
is faulty
Contact our custom-
er service