9
SAFETY INSTRUCTIONS
It is important that the following safety instructions are ob-
served to prevent risks to residents as well as carers, and to
avoid any damage to the bed:
The instruction manual must be read and observed before
using the care bed.
It is vital to observe the information given on the rating pla-
te! The information on the rating plate is explained in detail
on
of this instruction manual.
In the event of any faults or defects that could endanger a
person, the bed must not be used.
Electrically-adjustable care beds must only be operated by
the resident after instruction by trained staff.
Before the bed is used for the first time, the operator must
be satisfied that it is safe to use and in good condition.
The castors must always be placed in the braked position to
ensure the resident does not fall when getting into or out of
the bed.
The bed can be moved into various positions. When doing
so, take care to ensure no objects or parts of the body are
located in the adjustment area.
Only care staff may adjust the side rails. When adjusting the
sleeping surface position, take care to ensure the residents
do not come into contact with the side rails to avoid trap-
ping any part of the body.
The functionality of the side rails must be checked every day.
They must not bear any load of over 75 kg vertically or over
50 kg horizontally.
When using CPR (optional, mechanical emergency lowering
of the back rest), always additionally relieve the load on the
back rest by hand to prevent the back rest dropping in an
uncontrolled manner.
The AYLEEN care bed features a battery-operated emer-
gency mode. This allows the one-off lowering of the sleep-
ing surface in the event of a power cut. To ensure this func-
tionality, we recommend an annual test of the 9 V batteries
as part of the technical safety check and regular battery
replacement every two years or after each emergency low-
ering.
The hand controller functions can be locked or released on
the rear side using the key switch. Check that the locking
function has taken effect on the hand controller.
„HC-147 hand controller symbols“.
The drive system used must be operated using a VDE-ap-
proved power source – 100–240 V, 50/60 Hz mains socket.
The mains connection cable must also be protected by a
mechanical strain relief device. Nevertheless, take care to
ensure that the cable is not damaged by sharp edges, me-
chanical loads or pinch/shear points.