33
MAINTENANCE INTERVALS
Every two years and after every emergency lowering
§
Replace the two 9 V block batteries
§
If the 9 V block batteries are not replaced during the
maintenance interval or after use, the electric emergency
lowering function will cease to work. The care bed
equipped with a mechanically CPR can also be used ex-
tensively without the electric emergency lowering func-
tion and conforms in full to the standards and require-
ments of Regulation (EU) 2017 / 745. When not using
the function, we recommend removing the batteries to
prevent damage caused by leaked battery acid.
Annually
§
Inspection and maintenance
As required
§
Lubrication of mechanical components
§
Replacement of worn components if a defect occurs.
DELIVERY AND ASSEMBLY
Malsch GmbH care beds are generally delivered fully assembled,
or they are assembled on site by company technicians or con-
tractual partners.
Check the delivered bed against documentation for completeness
and conformity.
Any defects or damage must be pointed out to the freight com-
pany immediately and noted on the delivery document.
Signing of the delivery documents by both parties is obligatory
before commissioning.
If necessary, e.g. for maintenance, simple assembly procedures can
also be performed by professional authorised persons.
After maintenance and servicing work has been com-
pleted, check the functionality of electrical systems.
The manufacturer offers technician training for
maintenance and servicing work on their care beds.
A certificate obtained as part of this training au-
thorises the holder to carry out technical work in-
dependently on the beds.