MAINTENANCE
The manufacturer is only liable for the safety and reliability of
the product if it is serviced regularly and used in line with the
safety instructions. If any significant defects are found during
maintenance work which mean the safe operation of the product
cannot be guaranteed, the product must be taken out of use.
Maintenance work must be carried out at least once a year.
Any defects that impair the function and safety of
the care bed must be resolved before the bed is
used again and must be reported to the responsible
person.
Only original replacement parts from Malsch GmbH
are permitted to be used.
Service and maintenance tasks must not be conduct-
ed when the bed is occupied. The residents or the
care staff must not conduct maintenance tasks.
Procedure
1. Visual inspection
Check the welded structures for cracked weld seams and for
plastic deformation and wear. The welded structures include the
undercarriage and the sleeping surface with the moving interior
components. Also check that all screw joints are firmly connect-
ed.
2. Level of protection and functionality check of the side rails
During the functionality check, determine whether the side rails
can be easily locked in place and ensure that no impermissible
wear or deformation is visible.
The spacing requirements stated in IEC 60601-2-52:2009 /
AMD:2015 are used for protection. This is shown by the follow-
ing diagram and table.
Fig. 1 (dimensions of a single-piece side rail)