Mechanical back rest release / CPR
(optional)
Design with GS / design without side rails
Handle in the middle of the bed below the side panel
When operating the mechanical
release, hold the back rest in the
current position and, if possible,
relieve some of the load. Pull the
release lever towards the back
rest to disengage the back rest
and lower it manually into the end
position.
Press the S1 button on the hand controller again to reactivate
electrical back rest adjustment.
Caution! Before pressing the lever, ensure that
there are no objects or parts of the body below the
back rest. Manually relieve the weight on the back
rest during adjustment to prevent it dropping in an
uncontrolled manner.
DESCRIPTION OF FUNCTION
Depending on the design and type of care bed, it can be moved
into various positions by adjusting the back rest, the thigh rest, low-
er leg rest and the lift adjustment:
Back rest
Use the relevant buttons on the hand controller to adjust the
back rest.
(Back rest operating buttons)
When raised, the back rest can be adjusted by 120 mm over its
usual length to the head board (not with Editions 400 HC / 400 N).
This function (mattress compensation) allows the residents to sit
in a comfortable position while avoiding compression of the but-
tocks area and without compressing or restricting the stomach
or upper body.
Caution! The back rest is designed for the mechani-
cal load applied when raising a reclining person with
a maximum resident weight according to the speci-
fication on the rating plate. Sitting on the back rest
is not the intended specific function and may result
in damage and injury.