background image

24 

|

Mechanical back rest release / CPR

(optional) 

Design with GS / design without side rails

Handle in the middle of the bed below the side panel

When operating the mechanical 
release, hold the back rest in the 

current position and, if possible, 

relieve some of the load. Pull the 
release lever towards the back 
rest to disengage the back rest 
and lower it manually into the end 
position.

Press the S1 button on the hand controller again to reactivate 
electrical back rest adjustment. 

Caution! Before pressing the lever, ensure that 
there are no objects or parts of the body below the 
back rest. Manually relieve the weight on the back 
rest during adjustment to prevent it dropping in an 
uncontrolled manner.

DESCRIPTION OF FUNCTION

Depending on the design and type of care bed, it can be moved 

into various positions by adjusting the back rest, the thigh rest, low-

er leg rest and the lift adjustment: 

Back rest

Use the relevant buttons on the hand controller to adjust the 
back rest. 

 

 (Back rest operating buttons)

When raised, the back rest can be adjusted by 120 mm over its 
usual length to the head board (not with Editions 400 HC / 400 N). 

This function (mattress compensation) allows the residents to sit 
in a comfortable position while avoiding compression of the but-
tocks area and without compressing or restricting the stomach 
or upper body. 

Caution! The back rest is designed for the mechani-
cal load applied when raising a reclining person with 
a maximum resident weight according to the speci-
fication on the rating plate. Sitting on the back rest 
is not the intended specific function and may result 
in damage and injury.

Summary of Contents for AYLEEN CARE BED

Page 1: ...CARE DESIGN AYLEEN CARE BED INSTRUCTION MANUAL...

Page 2: ...2...

Page 3: ...re while maintaining the highest safety standards Thank you very much for the trust you have placed in us Our company guarantees carefully selected materials and contin uous quality control while empl...

Page 4: ...rcarriage Edition 400HC 20 AYLEEN care bed with undercarriage Edition 400N 21 AYLEEN care bed with undercarriage Edition XL 22 AYLEEN care bed Edition 500 23 Description of function 24 Back rest 24 Me...

Page 5: ...and control holder 36 Integrated bed linen holder 36 Troubleshooting 37 Product safety 38 Cleaning and disinfection 39 Disinfection 39 Care of wooden parts 39 Safe decommissioning storage 40 Warranty...

Page 6: ...RODUCT RANGE AYLEEN care bed with undercarriage Edition 400HC AYLEEN care bed with undercarriage Edition XL AYLEEN care bed with undercarriage Edition 400N AYLEEN care bed with undercarriage Edition 5...

Page 7: ...and care homes or compa rable medical facilities This complies with working environment 3 and 5 as stipulated by IEC 60601 2 52 2009 AMD1 2015 Prior written consent from Malsch GmbH is required if th...

Page 8: ...st time The instruction manual must be used for this training The instruction manual has been written for the AYLEEN care bed The images graphics and texts it contains may differ from the equipment su...

Page 9: ...e the residents do not come into contact with the side rails to avoid trap ping any part of the body The functionality of the side rails must be checked every day They must not bear any load of over 7...

Page 10: ...accordance with IEC 60601 2 52 2009 AMD 2015 when choosing a mattress it is important that there is a minimum distance of 22cm between the top of the sleep ing surface and the top of the side rail in...

Page 11: ...the manufacturer as part of the inspections and maintenance service for medical devices Electrically operated care beds are active medical devices and must be listed in an inventory for each site in...

Page 12: ...GmbH can cause increased emissions of electromagnetic interference or reduce the electromagnetic interference resistance of the device and thus lead to faulty operation Portable HF communication devic...

Page 13: ...13...

Page 14: ...ss com pensation AYLEEN care bed with undercarriage Edition 400N 100x206 90x200 27 to 81 Approx 120kg Total 180kg 145kg resident 20kg mattress 15kg accessories 33 71 12cm Mattress com pensation AYLEEN...

Page 15: ...4 Protection class II Lifting system force Lifting system force 1x6000N 2x6000N Sleeping surface adjustment force 2x4000 N 4x4000N Motor running time on 2 max off 18 min Data on operation transport an...

Page 16: ...t Operation Protection 1 2 3 4 5 6 7 8 9 10 11 12 NEXT I N S P E C T I O N 2024 100 240VAC50 60Hz 2 1 0 9A 35VDC2A max 2min ON 18min OFF IPX4 145kg 180kg 300kg 01 4065848000106 21 012104001234567 240...

Page 17: ...ical equipment from water spray only use in dry areas 7 Indicates the next technical check after delivery in line with VDE0751 1 8 Explanation of the safety symbols used on the rating plate Labelling...

Page 18: ...body weight and height are below these limits or who have a BMI under 17 as the risk of injury is increased for this group Symbol Label indicating replaceable mattresses in line with IEC 60601 2 52 20...

Page 19: ...igh rest down S6 Auto contour down S8 Height adjust ment down HC 147 HAND CONTROLLER SYMBOLS R1 LOCKED No functions operational R2 R4 UNLOCKED All functions available Locking functions positioned on t...

Page 20: ...or IV drip holder and accessories 4x surrounding the bed 7 Full length side rail 8 Release button for full length side rail 9 Lower leg rest adjustable ratchet 10 100mm single castors with castor brak...

Page 21: ...for IV drip holder and accessories 4x surrounding the bed 7 Full length side rail 8 Release button for full length side rail 9 Lower leg rest adjustable ratchet 10 100mm single castors with castor br...

Page 22: ...for IV drip holder and accessories 4x surrounding the bed 7 Full length side rail 8 Release button for full length side rail 9 100mm twin castors with castor brake1 AYLEEN care bed with undercarriage...

Page 23: ...1 2 1 Adjustable lower leg rest 2 Adjustable thigh rest 3 Static seat section 4 Adjustable back rest 5 Lower leg rest adjustable ratchet 6 Height adjustable feet FUNCTION ILLUSTRATION AYLEEN care bed...

Page 24: ...PTION OF FUNCTION Depending on the design and type of care bed it can be moved into various positions by adjusting the back rest the thigh rest low er leg rest and the lift adjustment Back rest Use th...

Page 25: ...ontour Ensure that that no ob jects or parts of the body are located in the area of the lifting mechanism Height adjustment Use the relevant buttons on the hand controller to adjust the height Height...

Page 26: ...bed with undercar riage Edition 500 is a stationary lifting frame to retrofit existing bed conver sions To do so the lifting frame is in stalled in place of the slat frame The horizontal adjustment wi...

Page 27: ...g when using the inser tion frame in a double bed Before adjusting the sleeping surface it is imperative to check that no body parts are below the moving elements Caution Prior to the first use of the...

Page 28: ...ion Slightly lift the upper side rail bar by the side rail groove B while at the same time pushing in the safety button A Slowly lower the side rail to the lowest position Repeat the process at the ot...

Page 29: ...for standard sleeping surface 90x200cm dimensions for other lengths will differ When using the comfort sleeping surface DS side rail dimensions When using with ABS metal mesh metal slats and wooden sl...

Page 30: ...e and maintenance tasks must not be conduct ed when the bed is occupied The residents or the care staff must not conduct maintenance tasks Procedure 1 Visual inspection Check the welded structures for...

Page 31: ...sections D 60 or D 318 E Distance between split side rails with reclining surface in a level position E 60 or E 318 F The largest dimension in at least one direc tion of every opening below the side r...

Page 32: ...h specifications at least every 2 years or after every case of emergency lowering 9 Measurements in line with DIN EN 62353 Check the electrical components of the care bed as specified in DIN EN 62353...

Page 33: ...a defect occurs DELIVERY AND ASSEMBLY Malsch GmbH care beds are generally delivered fully assembled or they are assembled on site by company technicians or con tractual partners Check the delivered b...

Page 34: ...of the components being disposed of are infectious contaminated If the bed is scrapped the wood plastic and metal parts used in its construction must be separated and disposed of properly If you have...

Page 35: ...g surface frame to the left or right of the head foot end The IV drip holder is only intended for attaching IV drips and not for hanging up other accessories or similar objects The maximum load is 8kg...

Page 36: ...reach of the resident Caution The hand control holder is flexible and must not be used as a standing aid or as a grab handle Integrated bed linen holder The integrated bed linen holder can be extended...

Page 37: ...safe working load has been exceeded Reduce the load Drives cut out after a long period of oper ation The adjustment time or safe working load has been exceeded and the control unit has reacted to ove...

Page 38: ...edical electrical equipment General requirements for safety DIN EN 60601 1 2 Medical electrical equipment Electromagnetic EM disturbances IEC 60601 1 6 Medical electrical equipment Usability DIN EN IS...

Page 39: ...d and protect the drive system from moisture We provide separate instructions for cleaning and disinfecting our care beds Care of wooden parts Malsch care beds only use wood surrounds that are finishe...

Page 40: ...care beds so that the wood surround and the frame cannot be damaged To ensure the necessary regular maintenance intervals attach the storage date on the bed in a visible location Lock the hand control...

Page 41: ...915 100 Fax 49 0 6626 915 116 SRN DE MF 000005173 declares under its sole responsibility that the devices named below comply with the essential requirements and provisions of Regulation EU 2017 745 fo...

Page 42: ...anagement system specified above and is monitored regularly The current validity is verifiable at www proficert com Original certificates contain a glued hologram T V Technische berwachung Hessen GmbH...

Page 43: ...43 NOTES...

Page 44: ...03 0 EN Updated 01 2023 Rev 3 0 Colours may vary Subject to technical changes Malsch GmbH Rohbergstra e 9 36208 Wildeck Obersuhl Germany Tel 49 0 6626 915 100 Fax 49 0 6626 915 116 bettenmalsch com in...