C .2-a Programmable Mode Setting
User can set up the parameters of up to 10 programs in Programmable Mode. The
setting concept of each program is consistent with the setting on the device.
*Note (11):
If user doesn’t need all the loops in the program, set the cycle at 0 in the next loop. For
example, only 3 loops are needed in the program; then just set 0 in the column of cycle in
Loop4, the system will automatically ignore the rest settings of loops. But each program
should contain at least 1 loop. In that, the cycle of
the first loop
in every program is
restricted to the range from 1 to 999 times. And the cycle of the rest loops is from 0 to
999.
*Note (12):
If user only needs 1 step or 2 steps in a loop, set 0 at both of the time columns (hour and
minute). The system will ignore the rest of steps in the same loop. For instance, the time
columns are both set at 0 in Loop1, Step2; and then the program will run Loop1, Step1, and
skip the rest steps in Loop1, directly proceed to Loop2.
*
Notice:
Remember to click
“Upload”
every time user finishes setting up or changing the
parameters. If user misses out this step, the settings will remain in the previous.
Edit the column to input user’s desired program
name. User can type 8 letters, including English
letters, numbers and punctuations.
The first program.
Select the number of loop to set up
the related parameters. Each
program contains 9 loops
*
11
.
Repetition of each loop. User can
set up the cycle up to 999 times.
Select the number of step to set up
the related parameters. Each loop
contains 3 steps
*
12
.
The running time for the assigned step. The
maximum is 999 hours and 59 minutes.
Input the values of hour and minute in the
respective columns.
The shaking speed set for the
assigned step. The maximum is
500 rpm.
The incubator temperature set for
the assigned step. The setting range
is from 15
℃
to 65
℃
.
Summary of Contents for Winpact SI-200
Page 2: ...2...
Page 14: ...14 PT 100 Sensor Port External Temp Port Spare Installation Hole Fan...
Page 27: ...27 A maximum of 9 microplate holders can be installed if the same type of holders is used...
Page 37: ...37 Features of Control Panel LCD Screen Key Pad 1 2 3 5 4 6 7 8...
Page 61: ...61 Step8 When the installation complete press Close to exit the installation dialogue...