43
Switch to “
step
” column to set up all the parameters in each step. Select the step by
pressing
and
. Push
and then
to set up the running time.
*
5
Use
and
to adjust the numbers; then press
or
to move from the hour
and the minute parameter. After finishing the time setting, push
to confirm the
adjustment. Then use
to move on to the rest parameters. Press
to start
setting up and press again to confirm the adjustment.
*Note (3):
The program will run the loops in sequence from Loop1 (Step1-Step3) to Loop9
(Step1-Step3).
*Note (4):
If user doesn’t need all the loops in the program, set the cycle at 0 in the next loop. For
example, only 3 loops are needed in the program; then just set 0 in the column of cycle in
Loop4, the system will automatically ignore the rest settings of loops. But each program
should contain at least 1 loop. In that, the cycle of
the first loop
in every program is
restricted to the range from 1 to 999 times. And the cycle of the rest loops is from 0 to
999.
*Note (5):
If user only needs 1 step or 2 steps in a loop, set 0 : 00 in the time column. The system will
ignore the rest of steps in the same loop. For instance, the time is set at 0 : 00 in Loop1,
Step2; then the program will run Loop1, Step1, and skip the rest steps in Loop1, and
directly proceed to Loop2.
Summary of Contents for Winpact SI-200
Page 2: ...2...
Page 14: ...14 PT 100 Sensor Port External Temp Port Spare Installation Hole Fan...
Page 27: ...27 A maximum of 9 microplate holders can be installed if the same type of holders is used...
Page 37: ...37 Features of Control Panel LCD Screen Key Pad 1 2 3 5 4 6 7 8...
Page 61: ...61 Step8 When the installation complete press Close to exit the installation dialogue...