33
For the flask of 1000 and 2000 ml, it is suggested to have spring band which is included with the
purchase of 1000/2000 ml flasks to prevent any leakage or over deviation while device is under
operation, especially at the high speed. For the rest of types of flasks, it is suggested to have a cable tie
to make sure flasks are firmly fixed with flask holder on the plate. (Refer to the pictures below)
Figure of the demonstration of the spring
band position on flask holder (for flask type
above 1000 ml)
Spring band
Figure of the demonstration of
the cable tie position on flask
holder (for flask type below 1000
ml)
Cable tie
NOTE:
-
Stop operation and check the tightness of
flask holders, spring band etc., and the
evenness of loading on platform, the speed
limit when there is any abnormal noise of
collision of flasks or any other noise while
operation.
NOTE:
The rubber stopper of flask holder can
be removed if handled incautiously, be
aware of them and check before start
operation each time.